Friction-aire
Typical
operating characteristics |
Temperature range:
+ 25 C above ambient to +150 C (302 F)
Typical air temperature uniformity:
at + 100 C: 0.5 C
Run-up time: (no load)
to + 150 C: 120 minutes
Cool downtime: (no load)
to +50 C: 30 minutes
Control accuracy: 0.5 C
Drift: (over 24-hour period)
at +50C: 1 C
at + 100C: 0.75 C
at + 150 C: 0 50 C
Average velocity: (adjustable)
Maximum: 400/FPM
Minimum: 100/FPM |
These three models have been listed by Underwriter's Laboratories for operation in Class
1, Group D hazardous locations. They are ideally suited for the testing and processing of
hazardous materials, paints, solvents and lacquers. There are no heating elements in the
chamber and no inherent hot spots to create potential danger. Tests show that load
temperature follows closely behind chamber temperature. If the application calls for the
processing of hazardous materials over extended periods, the load temperature and chamber
temperature will be practically equal all of the time. When chamber and load reach
complete equilibrium they will remain so indefinitely until the temperature is
deliberately changed. Even if spillage of flammables inside the chamber occurs, danger is
minimized since any surface the liquid encounters is only as hot as the interior
environment.
| Model |
Inside Dimensions
inches (cm)
W x D x H |
Overall Dimensions
inches (cm)
W x D x H |
HP |
1Approx Heating Power (Watts) |
Cubic Feet
Capacity |
Shipping Weight
Lbs. (Kg) |
| HS-1002 |
14 x 17 x 12
(35)x(43)x(31) |
37 x 33 x 48
(93)x(84)x(120) |
1.0 |
500 |
1.3 |
465
(211) |
| HS-1202 |
25 x 20 x 20
(63)x(52)x(51) |
42 x 39 x 63
(106)x(98)x(159) |
2.0 |
1300 |
5.2 |
800
(363) |
| HS-3802 |
38 x 25 x 25
(95)x(63)x(63) |
57 x 44 x 68
(144)x(111)x(172) |
3.0 |
2000 |
13.7 |
1300
(590) |
1Voltage
- all models: 208/240/480 V 3-Phase / 60 Hz. AC Chambers will operate satisfactorily on
208 V
3-Phase / 60 Hz. AC when motor is connected for 240 V 3-Phase /60 Hz AC, 2
Bench Model. |
All
specifications are subject to change without notice |
Available Option: Electric OTP with Door
Switch. All electric - Class 1, Group D, non-indicating over-temperature system to act as
backup to main over-temperature control. Provides extra protection for critical loads.
Note: This option deletes UL listing of unit.
36 Cubic foot model for Class I, Group D
applications
This model offers all of the operational advantages of the Friction-Aire
in a size that meets higher volume processing requirements. Its design resembles the
typical batch oven configuration and provides a work chamber volume of 36 cubic feet. It
delivers controlled batching temperatures in a range from 25 C above ambient to + 150
C (320 F)
These ovens have been specially engineered to meet the stringent design requirements of
Class 1, Group D service. They incorporate a relief panel and safety cage centered in the
back wall, plus Class 1, Group D motors and starters with approved M.I. cable and
fittings. An electric non-indicating backup OTP in a Class 1, Group D enclosure is
furnished as standard on this unit. A Class 1, Group D door switch that shuts off the
blower when the door is opened is also included as standard.
Typical
operating characteristics |
Temperature range:
+ 25 C above ambient to + 150 C (302 F)
Typical air temperature uniformity:
at + 100 C: + 0.5 C at + 150 C: 1.0 C
Typical maximum rate of rise: (no load)
to + 150 C: 60 minutes
Maximum rate of cooling: (no load)
to +50 C: 30 minute
Control accuracy: 0.5 C
Drift: (over 24-hour period)
at + 50 C: + I C
at + 100 C: 0.75 C
at + 150 C: 0.50 C
Average velocity: (adjustable)
Maximum: 500/FPM
Minimum: 200/FPM |
| Model |
Inside Dimensions
inches (cm)
W x D x H |
Overall Dimensions
inches (cm)
W x D x H |
HP |
*Approx. Heating Power (Watts) |
Cubic Feet Capacity |
Shipping Weight
Lbs. (Kg) |
| *HS-362-1 |
37 x 37 x 48
(93)x(93)x(120) |
66 x 58 x 91
(167)x(147)x(231) |
10.0 |
6500 |
36.0 |
2200
(1000) |
*Bench Model. * Voltage -all models:
208/240/480 V 3- Phase/ 60 Hz. AC Chambers will operate satisfactorily on 208
V 3-Phase
/ 60 Hz. AC when motor is connected for 240 V3-Phase /60 Hz. AC |
All
specifications are subject to change without notice |
For higher temperature
operations
These ovens are not UL Listed because of their increased temperature
capabilities. They exceed temperatures allowed in a Class 1, Group D location, but they
have the same basic construction and operating characteristics of the other HS models
previously described. An electric non-indicating backup OTP in Class 1, Group D enclosure
is furnished as standard on these higher temperature units. A Class L Group D door switch
that shuts off the blower when the door is opened is also included as standard.
| Model |
Inside Dimensions
inches (cm)
W x D x H |
Overall Dimensions
inches (cm)
W x D x H |
HP |
*Approx. Heating Power (Watts) |
Cubic Feet Capacity |
Shipping Weight
Lbs. (Kg) |
| HS-1004-1 |
14 x 17 x 12
(35) x (43) x (31) |
37 x 33 x 48
(93) x (84) x (120) |
1.0 |
600 |
1.3 |
465
(211) |
| HS-1204-1 |
25 x 20 x 20
(63) x (51) x (61) |
42 x 39 x 63
(106) x (98) x (159) |
3.0 |
2000 |
5.2 |
800
(363) |
| HS-3804-1 |
38 x 25 x 25
(95) x (63) x (63) |
57 x 44 x 68
(144) x (111) x (172) |
5.0 |
3000 |
13.7 |
1300
(590) |
1BenchModel. * Voltage
-all models: 208/240/480 V 3-Phase/ 60 Hz AC Chambers will operate satisfactorily on
208 V 3-Phase / 60 Hz AC when motor is connected for 240 V 3-Phase /60 Hz
AC |
All
specifications are subject to change without notice |
Typical
operating characteristics |
Temperature range:
+ 20 C above ambient to +260 C (500 F)Typical air temperature uniformity:
at + 100 C: 0.5 C
at + 200 C: 1.0 C
Typical maximum rate of rise: (no load)
to + 260 C: 140 minutes
Maximum rate of cooling: (no load)
to +50 C: 30 minutes
Control accuracy: 0.5 C
Drift: (over 24-hour period)
at +50 C: l C
at + 100 C: 0.75 C
at + 150 C: 0.50 C
at + 200 C: 0.50 C
Average velocity: (adjustable)
Maximum: 500/FPM
Minimum: 250/FPM |
|